Abstract

In this study, a feedback device of force and temperature sensations for myoelectric prosthetic hand users was developed. When a prosthetic hand user grasps an object using the myoelectric prosthetic hand, the stiffness and temperature of the object are measured using sensors attached to the prosthetic hand, and force and temperature sensations are fed back to the upper arm of the user. From the experimental evaluation of the feedback device, the influence of temperature change on force sensations was confirmed. Therefore, to feed back the same force sensation to the user even if a temperature change has occurred, compensation functions were derived using the maximum likelihood method. On the basis of paired comparison, verification experiments were conducted, which demonstrated the effectiveness of the derived compensation functions.

Content from these authors

This article cannot obtain the latest cited-by information.

© 2020 Fuji Technology Press Ltd.

This article is licensed under a Creative Commons [Attribution-NoDerivatives 4.0 International] license (https://creativecommons.org/licenses/by-nd/4.0/).
The journal is fully Open Access under Creative Commons licenses and all articles are free to access at JRM Official Site.
https://www.fujipress.jp/jrm/rb-about/
Previous article Next article
feedback
Top