Journal of Surface Analysis
Online ISSN : 1347-8400
Print ISSN : 1341-1756
ISSN-L : 1341-1756
Technical Report
Development of a sample transfer vessel by using a coaxial type linear motion feedthrough
E. KobayashiJ. MeikakuH. SetoyamaT. Okajima
Author information
JOURNAL FREE ACCESS

2012 Volume 19 Issue 1 Pages 2-5

Details
Abstract
We have developed a compact sample transfer vessel. It consists of coaxial type linear motion feedthrough, vacuum vessel and sample catcher. The sample can be analyzed by using this system without exposing to atmosphere. The vessel pressure reached ∼ 4.6 × 10-4 Pa and kept 1 × 10-3 Pa for 20 days. The aluminum on which an argon sputtering was carried out was kept in this equipment for 41 hours. Then, ∼ 1.0 nm thick oxide film was formed on the surfaces. The oxide film thickness was about 1 ⁄ 4 of the natural oxide layer thickness.
Content from these authors
© 2012 by The Surface Analysis Society of Japan
Next article
feedback