Journal of the Japan Society of Applied Electromagnetics and Mechanics
Online ISSN : 2187-9257
Print ISSN : 0919-4452
ISSN-L : 0919-4452
[Academic Paper]
Evaluation of radial stability of maglev blood pump for extracorporeal circulation driven to synchronize with heart beat
Ryota SATOHToru MASUZAWAMasahiro OSAKouhei OHNOTakashi NISHIMURAShunei KYO
Author information
JOURNAL FREE ACCESS

2018 Volume 26 Issue 2 Pages 362-367

Details
Abstract

 We have developed a novel blood pump with a magnetic levitated impeller to recover cardiac function. The blood pump can increase the coronary flow by varying rotational speed of the impeller in accordance with the heartbeat. In this work, radial displacements of the rotor were measured to evaluate passive stability with improved impeller which generated stronger axial attractive force. Levitation stability of the impellers in the pump with horizontally setting was compared to that vertically setting under the condition of continuous flow and pulsatile flow in a mock flow loop with normal saline. As a result, touchdown occurred with present impeller in the condition of vertical setting and pulsatile flow, whereas touchdown did not occur with improved impeller in any condition. In the future, hemolysis test, vibration test and durability test will be carried out with improved impeller. In consideration of safely, the levitation stability of improved impeller would be improved by changing size of the permanent magnets.

Content from these authors
© 2018 The Japan Society of Applied Electromagnetics and Mechanics
Previous article Next article
feedback