Journal of the Japan Society of Applied Electromagnetics and Mechanics
Online ISSN : 2187-9257
Print ISSN : 0919-4452
ISSN-L : 0919-4452
[Academic Papers]
Development of a Muscle-Equivalent Phantom for Electrical Impedance Tomography
Seiya TOYODATakahiko YAMAMOTO
Author information
JOURNAL FREE ACCESS

2022 Volume 30 Issue 2 Pages 144-149

Details
Abstract

 This paper describes the development of a muscle-equivalent phantom for electrical impedance tomography (EIT). Conventional phantom adding titanium dioxide coated with antimony-doped tin oxide (ATO/TiO2) cannot mimic the electrical properties of the human body over wide frequency band. The main material of the ATO/TiO2-added phantom was changed from water to glycerol, and polyethylene powder (PEP) was added to decrease the relative permittivity of the medium, in order to expand the frequency band that can be mimicked. The electrical properties were measured for varying additions of glycerol, PEP and ATO/TiO2, and the results agreed with the theoretical equations. Muscle-equivalent phantom for 10 kHz to 10 MHz were developed, and the errors in the electrical properties were 8.55% and −7.14% for relative permittivity and conductivity, respectively. The combination of a development phantom with a skin-equivalent phantom and a fat-equivalent phantom allows for more accurate evaluation of EIT devices.

Content from these authors
© 2022 The Japan Society of Applied Electromagnetics and Mechanics
Previous article Next article
feedback
Top