Journal of the Japan Society of Applied Electromagnetics and Mechanics
Online ISSN : 2187-9257
Print ISSN : 0919-4452
ISSN-L : 0919-4452
Special Issue on The 33th MAGDA Conference in TOKYO (MAGDA2024)
Analysis and Actual Measurement of Coil-to-Coil Capacitance of High-Frequency Isolation Transformers for Suppression of High-Frequency Patient Leakage Currents
Placing the Transformer inside an Aluminum Case
Saori NISHITANIKenji SHIBA
Author information
JOURNAL FREE ACCESS

2025 Volume 33 Issue 3 Pages 293-299

Details
Abstract
High-frequency isolation transformers are used to suppress high-frequency patient leakage current flowing in the insulated parts of medical devices. In this study, we assumed that the high-frequency isolation transformer would be placed in an aluminum case for actual case and researched whether the coil-to-coil capacitance was small enough to suppress the high-frequency patient leakage current below the regulated value. The coil-to-coil capacitance was analyzed and measured by changing the items laid in the isolation transformer, the presence of PTFE wrapped around the isolation transformer, and the number of ferrite cores used in the isolation transformer. The results of the analysis in the electromagnetic field simulation showed that the smallest capacitance was 5.040pF when two ferrite cores were used, PTFE tape was wound, and a 1.0mm PTFE plate was laid. The actual measurement also showed a minimum of 3.9807 pF in this case.
Content from these authors
© 2025 The Japan Society of Applied Electromagnetics and Mechanics
Previous article Next article
feedback
Top