Artificial Pancreas
-Clinical application of the new glucose sensor-
K. ICHIKAWAT. YUKAWAK. OOKURAH. ICHIHASHIT. KONDOH. KOJIMAY. SAWADAS. IKEDAK. ITOC. YAMAZAKIK. MASUKO
Author information
JOURNAL FREE ACCESS

1981 Volume 10 Issue 6 Pages 1105-1108

Details
Abstract
The vessel access type glucose sensor for the implantable artificial pancreas has been developed in our labolatories. The sensor is connected to a vessel access through which the blood flows and enable to measure blood glucose levels without sampling and diluting blood.
Clinically, the sensor was applied on renal failure patients using hemodialysis circuit. The sensor responded well in parallel with the blood glucose levels measured by Autoanalyser from 110 to 350mg/dl in increasing phase, however, at initial measurement it takes about 30 minutes' delay to response in decreasing phase.