Study of the dialyser performance on alteration of the outlets and inlets of dialysate
A. KANAMOTOK. KATAM. YONEMITSUH. ANDOH. IKEBUKUROK. SATOM. NOZAWA
Author information
JOURNAL FREE ACCESS

1982 Volume 11 Issue 6 Pages 1253-1256

Details
Abstract
The new modified HFK has three pairs of dialysate connectors on the side of HFK housing symmetrically so that we are able to change its effective surface area and dialysate flow length easily and repidly. In vitro studies were made with available nine ways of dialysate flow by utilizing two dialysate connectors (others should be closed with caps). VB 12 and Chrolide clearance were measured and compared respectively. Conclusion: In practise, modified HFK should have one dialysate inlet on the lower part of housing and three outlets on the opposite side so that the values of clearance are controlled easily.