Plasticizers released from the extracorporeal circulation devices have been suspected harmful
I. NAKAGAWA[in Japanese][in Japanese][in Japanese][in Japanese][in Japanese][in Japanese][in Japanese][in Japanese][in Japanese][in Japanese][in Japanese]
Author information
JOURNAL FREE ACCESS

1985 Volume 14 Issue 1 Pages 34-36

Details
Abstract
For the attempt at preventation from the plasticizers leak, inner surfaces of the blood tubings were coated by thin layer of gelatin or urethane. Plasma levels of di-ethylhexyl phthalate (DEHP) were measured during the hemodialysis treatment using the 2 types of coated blood tubings. Increments of DEHP levels were significantly suppressed by both gelatin and urethane coating. Gelatin decreased especially the plasma DEHP levels by 60%. Our results indicated that both gelatin and urethane coating of the blood tubings were effective to prevent from the release of plasticizers.