Selection of the artificial valves for ventricular assist device (VAD) from the view point of hydrodynamic behaviors of the valves
M. UMEZUT. TANAKAH. TAKANOT. NAGATAK. TSUCHIYAT. AKUTSU
Author information
JOURNAL FREE ACCESS

1986 Volume 15 Issue 2 Pages 613-616

Details
Abstract
Hydrodynamic data, especially the water hammer (WH), for five valves (four disk and one polyurethane) were obtained to select appropriate valves for VAD pump. The results were as follows: 1) WHs by Björk-Shiley monostrut (BS(MS)) were higher than those by BS. When BS (MS) was used as an inlet valve, irregular behavior of the disk was observed at low %-systole conditions. 2) When the disk material was changed from pyrolytic carbon to delrin, WH decreased. 3) WHs by Medtronic-Hall were slightly higher than those by BS. 4) Conduit type tri-leaflet polyurethane valves developed in our lab for VAD use, exhibited low WH as compared to disk valves.