1991 Volume 20 Issue 3 Pages 1138-1143

Details
Abstract
During clinical hypothermic cardiopulmonary bypass, performances of three groups of membrane oxgenators (Bard HF-5000(H), MERA EXCELLING α(M), Capiox-E (C)) have been studied in 39 patients with aortocoronary bypass operation for ischemic heart disease. (1) O2transfer performance was more excellent in (H) than in (M) and (C). (2) Plasma free hemoglobin was higher in (C) group than (H) and (M) group. (3) Platelet count postoperatively was higher in (H) group than in (M) and (C) group. (4) Coagulation factors were not significantly different in each three group.