MAGNETICALLY ACTUATED LEFT VENTRICULAR ASSIST DEVICE
H. TANAKAM. FUNAMIT. TAKABAS.g. KOVACSL.e. ONDROVICM. AIBAP.p. MCKEOWNR.g. CONNAR
Author information
JOURNAL FREE ACCESS

1991 Volume 20 Issue 3 Pages 663-665

Details
Abstract
A magnetically actuated left ventricular assist device which has small size and simple mechanism has been developed. This device consists of 4-main subassemblies. 1) actuator (size 32.5cm3, weight 312g). 2) magnetic pusher-plate diaphragms. 3) pump-chamber with valve parts. 4) interface mount. Pumping force is generated by an electromagnetic actuator whose (magnetic flux) core is centered on the axis of the rare-earth permanent magnet, which in turn is centrally mounted on the pump pusher-plate diaphragm. This device's hydraulic power is 7.8- 11.2 watts. Using Mock loop model, this pump generates 3.2-5.8 1/min out put with 120mmHg afterload and 8-12mmHg preload when heart rate was changed from 80 to 120 bpm.