Abstract
We have studied three-dimensionally (3-D) reconstructed images of atrioventricular (AV) annuli based on four-chamber magnetic resonance (MR) images for total artificial heart (TAH) implantation. To assess the interrelationship among the AV annuli and the thorax, we applied the frame of reference of the vertebral canal to these images. The Z-axis was along the vertebral canal, the Y-axis in the anteroposterior direction. The angle alpha was defmed as the angle between the X-axis and the intersection between the plane of the annulus and the X-Y plane. The angle beta was defined as the angle between the Z-axis and the intersection between the plane of annulus and X-Z plane. The angle gamma was defined as the angle between the plane of the annulus and the anterior thoracic wall. We determined each angle for the mitral annulus (MVA) and tricuspid annulus (TVA) in four normal subjects to be as follows: alpha-MVA, 20.5-393 degrees;alpha-WA, 26.1-43.5 degrees;beta-MVA, 4.7-49.4 degrees;beta-WA, 4.4-40.9 degrees;and gamma-WA, 35.2-44.1 degrees.
It is suggested that reconstruction of the thoracic wall and AV annuli using the frame of reference of the diaphragmatic surface of the heart would facilitate evaluation of the spacial requirements for TAH implantation. Research to define this new frame of reference is in progress.