Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
UTOMATED FABRICATION OF POLYURETHANE HEART VALVES
K. UCHIYAMAN. FUJISAWAY. SHIMAZAKIM. RANAWAKET. FUJIMOTOM. UMEZU
Author information
JOURNAL FREE ACCESS

1994 Volume 23 Issue 1 Pages 124-129

Details
Abstract
We have developed an automated “Arm Robot” Fabrication System (ARFS) to make the leaflets of a polyurethane (PU) valve for use in ventricular assist devices (FAD). Preliminary testing using the ARFS resulted in a 27% success rate of the valves fabricated, as compared to 50% for the hand made valves. To gain a better understanding of the deviations in thickness of the leaflet, we measured it at a 135 different points using a Miniature moire measuring system. Of the ARFS made valves, the max. deviation between the three leaflets was at 25um. and 47um. for the hand made valves, while the leaflet itself became thicker towards the base. This result indicated a much more even thickness distribution in the ARFS made valves. From valve fatigue testing and using the FEM (finite element method) it was evident a thickness of approximately 300um. at the base of the leaflet is necessary to achieve long term durability. Although problems remain the ARFS, as the thickness of the valve is better controlled using this system, it is reasonable to assume parameters of valve performance including, durability, leakage & pressure drop are also better controlled.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback