Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
AN ELECTROSTATIC LINEAR ACTUATOR DEVELOPED FOR AN ARTIFICIAL CARDIAC MUSCLE
-IMPROVEMENT IN PERFORMANCE OF THE ACTUATOR BY USING LIQUID CRYSTAL AS GAP -FILLING MEDIUM-
N. KABEIK. TSUCHIYA
Author information
JOURNAL FREE ACCESS

1995 Volume 24 Issue 2 Pages 345-350

Details
Abstract
We developed an electrostatic lineer actuator for use as an artificial cardiac muscle which consists of parallel plate capacitors and a slider made of dielectric material. The most effective methods to increase the output force of the actuator are optimizing the dielectric constant of the gap-filling medium, which fills the space between two electrodes of the capacitor, and increasing the dielectric strength of the medium for supplying high voltage to the capacitor. We chose liquid crystal as the gap-filling medium. The maximum velocity of the slider of the actuator was 2.24mm/s. The maximum output force of 56.5mN was generated by the actuator filled with liquid crystal. This value is seven times as large as that generated by the same actuator filled with pure water used as the gap-filling medium in previous experiments.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback