IMPLANTATION OF THE THIRD GENERATION IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (VENTAK PRx II) WITH THE SINGLE LEAD SYSTEM
H. HAKANOA. SHIIKAWAT. AKIMOTOK. HAYASHIM. HACHIDAH. NISHIDAM. ENDOH. KOYANAGI
Author information
JOURNAL FREE ACCESS

1995 Volume 24 Issue 2 Pages 495-498

Details
Abstract
Seven VT/VF patients were safely implanted the 3rd generation implantable cardioverter defibrillator (I CD: Ventak PRxII) with the single lead (SL) system. 2 patients were used subcutaneous leads (array lead: AL) together. The mean defibrillation threshold was 15.0+/-3.5 J in the group using only SL and was not significantly different from that of the group using AL together (22.5+/-3.5J). However, the mean impedance at each defibrillation testing in the AL group was apparently lower (30.1+/-6.9ohms) than that of the SL group (46.9+/-4.0ohms)(p<0.05). We concluded that the AL using with the SL was useful to prevent a loss of defibrillation energy.