CLINICAL EVALUATION OF INFANT MEMBRANE OXYGENATORS, “M-16” IN COMPARISON WITH “MINIMAX” AND “MENOX AL-2000”
T. SAKURADAA. YAMADAS. INOUEH. NAKAKURAT. MATSUZAKIT. UMAMIN. KAWAHARADAH. AJIKIS. KOMATSU
Author information
JOURNAL FREE ACCESS

1995 Volume 24 Issue 2 Pages 600-603

Details
Abstract
In open heart surgery for infants, we studied clinical applicability of the newly developed membrane oxygenator M-16 in comparison with the MINIMAX and the MENOX AL-2000. The M-16 has a special structure of rigid metal plates covered with silicone membranes and showed an excellent O2 and CO2 transfer performance compared with the other oxygenators. There were no statistical differences in hemocompatibility among the three types of oxygenator in terms of plasma free hemoglobin and platelet preservation ratio. But it was very difficult and complex to evacuate the air :frgm the :M-16 during the priming period, and the static priming volume (200ml) of this model is iativeyly miidh for infants.