RELATIONSHIP BETWEEN HUMAN BODY SIZE AND PROPER ARTIFICIAL VALVE SIZE IN THE AORTIC POSITION
K. KAWACHIK. MIZUGUCHIT. KAWATAS. KOBAYASHIY. HAMADAK. NIWAYAJ. HASEGAWAS. KITAMURA
Author information
JOURNAL FREE ACCESS

1996 Volume 25 Issue 3 Pages 586-589

Details
Abstract
In vivo valve function of the artificial valve in patients with St. Jude Medical (SJM) valve and Bicer prosthesis in the aortic position was evaluated both at rest and during exercise after surgery. Effective orifice area was obtained with use of the Gorlin formula. Transaortic pressure gradient was obtained from Gorlin formula in each body surface area.
For 21mm SJM valve, transaortic pressure gradient with body surface area 2.0m2 was 9mmHg at rest and 19mmHg during exercise. For 21mm Bicer valve, transaortic pressure gradient with body surface area 2.0m2 was 14mmHg at rest and 19mmHg during exercise. From these results it is considered that SJM valve and Bicer valve for the 21mm valve are available in the aortic position.