Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
In Vivo Study of the Performance of a Mechanical Seal Involving a Recirculating Cooling Water System Used in a Centrifugal Blood Pump
J TomiokaT MoriK YamazakiT AkimotoH Koyanagi
Author information
JOURNAL FREE ACCESS

2000 Volume 29 Issue 1 Pages 47-50

Details
Abstract
The integrity of the shaft seal is a critical factor in the success of an implantable centrifugal blood pump. A precision mechanical seal has been developed to address this requirement. The seal incorporates a recirculating cooling water system to provide lubrication and prevent deposition of blood elements. The objective of the current study was to experimentally evaluate the sealing properties of this mechanical seal in vivo. The leakage rate of plasma into the cooling water through the mechanical seal was measured by monitoring the infiltration of sodium ions from bovine blood. Measurement of the concentration of sodium ions was performed using inductively coupled plasma light emission analysis. The leakage rate of the cooling water into the blood through the mechanical seal was measured by monitoring the volume of the cooling water in the reservoir. Over a period of 153 days, the leakage rate of plasma was 0.022ml/day and the leakage rate of the cooling water was 0.497ml/day. In conclusion, the mechanical seal performed acceptably well to support its use in a centrifugal blood pump for chronic application.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback