Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
P Wave Amplitude Histogram Recorded in Patients with Implanted Single Pass Lead VDD Pacemakers.
H OgawaT IshikawaS SumitaI KobayashiT SuganoT ShigemasaT EndoK KimuraS Umemura
Author information
JOURNAL FREE ACCESS

2000 Volume 29 Issue 2 Pages 394-397

Details
Abstract
We studied P wave amplitude (PWA) histogram in 6 patients with implanted single pass lead VDD pacemakers.The pacemakers detected PWA automatically at 6-minute intervals and data were expressed on histograms by eight bins. The average age of patients was 62.3±17.8 years (mean±SD). The half-amplitudes of the atrial sensing threshold measured by conventional method correlated positively to those assessed by PWA histogram (r=0.77, p<0.0001). Incidences of atrial sensing failure were documented by the records of PWA histograms. In conclusion, PWA histogram is useful to determine adequate atrial sensitivity in patients with implanted single pass lead VDD pacemakers.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top