Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Multi-mode Artificial Heart Driving System
Makoto TAKAHASHIYoshinori MITAMURAKatsuyuki YAMAMOTOTomohisa MIKAMITakao NAKAMURAKiyotaka NISHIURAKazuhito SAKAKIBARA
Author information
JOURNAL FREE ACCESS

1978 Volume 7 Issue 1 Pages 141-144

Details
Abstract
Highly reliable pneumatic artificial heart (AB) driving system with complete redundant system for failure free operation was developed and used in 13 left ventricular (LV) bypass experiments without any failure. It has the following features besides normal counterpulsation; (1) When the contraction of LV is detected during pump systole, AH is quickly switched to diastolic phase. In tachycardia, if systolic time of AH becomes too short to eject enough blood flow, the next ejection begins earlier and continues longer. (2) When the diastole of LV continues for longer than 1+ seconds, AH starts the pumping. (3) In fibrillation, AH is asynchronously pumped.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback