Journal of Japan Society of Computer Aided Surgery
Online ISSN : 1884-5770
Print ISSN : 1344-9486
ISSN-L : 1344-9486
Originals
Surface contact control of ultrasonic therapy device on patient body surface using parallel-link robot
Kaoru NatsumeYoshihisa WadaShinya OnogiTakashi MochizukiKohji Masuda
Author information
JOURNAL FREE ACCESS

2016 Volume 17 Issue 4 Pages 343-354

Details
Abstract
For ultrasound therapy, e.g. microbubbles delivery, physical contact between therapeutic device and body surface is important. To guarantee the entire contact of the device, we propose a control method considering contact force and reaction moment using a robotic system and sensors. We used a parallel link robot, which has three links and the holder to determine position and angle of the device. First we established a series of surface contact controls, which consists of (a) initial contact control, (b) balance contact control, and (c) angle restoration control. Second we eliminated self-weight moment of the device using the sensors. Next we proposed a measurement method of elasticity coefficient of soft material. Then we performed evaluation experiments using soft material and human body. In the experiment using soft material, contacted area was visualized and evaluated with color ink. In the experiment using human body, convergence time to the desired position and angle was also evaluated while monitoring contact force and moment. We concluded that the determination of appropriate parameters is important and dominant for the quality of the system.
Content from these authors
© 2016 The Japan Society of Computer Aided Surgery
Previous article Next article
feedback
Top