Efficacy of Buttonhole Puncture for Arteriovenous Grafts
Kazutoshi Ogawa, Yasuo Mishima, Naohiro Hayashi, Ryoichi Tamaya, Yuhei Fujisawa, Yasushi Kakuchi
Author information
Keywords: graft, buttonhole, puncture
JOURNAL FREE ACCESS

2022 Volume 55 Issue 3 Pages 209-213

Details
Abstract

As the number of patients who have difficulty with arteriovenous fistulas is increasing, the use of arteriovenous grafts is rising. Although some reports have suggested that buttonhole puncture is effective for maintaining arteriovenous grafts, it is not recommended due to the risk of infection. We have prepared a buttonhole puncture manual, which recommends the use of dull needles to maintain arteriovenous grafts and limited punctures, and we have not experienced any vascular access infections in the 6 patients with arteriovenous grafts treated at our hospital over the past 7 years. As for puncturing, we measured the number of attempts and the success rate of buttonhole puncture for one month and found that the success rate was 97.9% and 50% when ≤10 and >10 punctures were performed, respectively. By strictly following the manual, the long‒term maintenance and preservation of arteriovenous grafts was possible.

Content from these authors
© 2022 The Japanese Society for Dialysis Therapy