Nihon Toseki Igakkai Zasshi
Online ISSN : 1883-082X
Print ISSN : 1340-3451
ISSN-L : 1340-3451
Evaluation of creatinine determination by enzymatic assay and the Jaffe reaction method in peritoneal equilibration test
Shinya NakamuraKaori TomonagaMidori HojoKikuo IitakaAkira Nishida
Author information
Keywords: CAPD
JOURNAL FREE ACCESS

1995 Volume 28 Issue 10 Pages 1375-1378

Details
Abstract
We measured creatinine in peritoneal dialysate by enzymatic assay (EA) and the Jaffe reaction method (J), to perform peritoneal equilibration tests (PET) in children. Average D2/P ratios measured by J (corrected by glucose), and EA of D2/P were 0.494 and 0.478 (p<0.01), respectively, and those of D4/P were 0.788 and 0.722 (p<0.01), respectively. EA is more accurate and useful for measuring creatinine in peritoneal dialysate since this method is not influenced by the glucose concentration of the dialysate.
Content from these authors
© The Japanese Society for Dialysis Therapy
Previous article Next article
feedback
Top