Original articles
The temperature of aural air stimulation equivalent to aural stimulation with water at 20°C in the caloric test.
Kazuhiko TaharaKazunori SekineGo SatoKazunori MatsudaSeiichiro KamimuraTakahiro AzumaNoriaki Takeda
Author information
JOURNAL FREE ACCESS

2015 Volume 74 Issue 1 Pages 15-19

Details
Abstract
 In the present study, we examined the temperature of aural air stimulation that was equivalent to aural stimulation with water at 20°C in the caloric test. In 10 ears of 5 healthy volunteers, the maximum slow phase eye velocities (MSPEVs) of nystagmus induced by aural stimulation with air at 22°C, 46°C and 16°C were the same as those with water at 30°C, 44°C and 20°C, respectively. These findings suggest that caloric stimulation with air at 16°C is equivalent to that with water at 20°C that is commonly used in the caloric test in Japan. The MSPEVs of nystagmus induced by aural stimulation with air at 16°C were over 20°/sec in all ears except one, in which the MSPEV was 19.7°/sec. The criteria of canal paresis where the MSPEV of caloric nystagmus induced by aural stimulation with water at 20°C is less than 20°/sec can be used in the caloric test with aural air stimulation at 16°C.
Content from these authors
© 2015 Japan Society for Equilibrium Research