Journal of the Japanese Society of Intensive Care Medicine
Online ISSN : 1882-966X
Print ISSN : 1340-7988
ISSN-L : 1340-7988
Efficacy and reliability of PEEP valves
Keita SetojimaKoichi TanigawaMakoto OokuboEiichi GotoKeiichi Tanaka
Author information
JOURNAL FREE ACCESS

2003 Volume 10 Issue 4 Pages 365-368

Details
Abstract
There are few data available on the efficacy and reliability of PEEP valves for a transport ventilator. We evaluated PEEP valves manufactured by Ambu(R), Inspiron(R), Smiths Industries(R), and Vital signs(R) co. The Smiths Industries(R) products had the biggest differences between the preset indicators and measured pressures at PEEP 5, 10, 15cmH2O in comparison with other products. Pressure decrements at PEEP 5, 10, 15cmH2O were larger in the Inspiron(R) and the Vital signs(R) products than in others, and the decrements tended to increase at higher PEEPs. Resistance created by the PEEP valves were negligible. We conclude that pressure monitoring is required for the safe and effective use of commercially available PEEP valves.
Content from these authors
© The Japanese Society of Intensive Care Medicine
Previous article Next article
feedback