Japanese Journal of Joint Diseases
Online ISSN : 1884-9067
Print ISSN : 1883-2873
ISSN-L : 1883-2873
Original Articles
Evaluation of the Accuracy of Distal Femoral Cutting Using a Portable Navigation Device in Total Knee Arthroplasty
Takahiro NOGUCHI[in Japanese][in Japanese][in Japanese][in Japanese]
Author information
JOURNAL FREE ACCESS

2016 Volume 35 Issue 1 Pages 17-23

Details
Abstract
Objective: Computer-assisted surgery systems improve alignment accuracy in total knee arthroplasty (TKA) but they are expensive and time-consuming and thus, unpopular. Recently, a portable navigation system has shown excellent performance. We examined the accuracy of distal femoral cutting using an accelerometer-based portable navigation device (KneeAlign® 2; OrthAlign Inc., Aliso Viejo, CA, USA) during TKA.
Methods: We included patients consisting of 78 knees with osteoarthritis, who had undergone TKA using the portable navigation KneeAlign® 2 system. Computed tomography for each lower extremity was performed preoperatively. We simulated the TKA components’ position by using the preoperative planning software Athena® (Soft Cube, Osaka, Japan), and we measured the flexion angle of the femoral component relative to the femoral mechanical axis in the sagittal plane. We performed osteotomy of the distal femur perpendicular to the femoral mechanical axis in the coronal plane and osteotomy of the distal femur using the measured flexion angle of the femoral component relative to the femoral mechanical axis in the sagittal plane.
Results: In all, 93.6% knees had a femoral component alignment within 3°perpendicular to the femoral mechanical axis.
Conclusions: The femur and tibia should be cut perpendicular to the mechanical axis on the coronal plane in TKA. Varus-valgus alignment of more or equal to 3°increased the incidence of component loosening. The portable navigation device ensured an accurate performance of femoral osteotomy in the coronal plane as that performed using other navigation systems. With regard to sagittal plane alignment, the flexion angle between the distal femoral anatomical axis and the mechanical axis showed large interindividual variations and this was measured beforehand using Athena®, and which was confirmed intraoperatively. However, this alignment was primarily achieved visually using the anterior femoral cortex intraoperatively via the conventional method without using a navigation system. Thus, a portable navigation device is useful when performing surgery according to a predetermined plan.
Content from these authors
© 2016 Japanese Society for Joint Diseases
Previous article Next article
feedback
Top