JSME International Journal Series B Fluids and Thermal Engineering
Online ISSN : 1347-5371
Print ISSN : 1340-8054
ISSN-L : 1340-8054
Investigation of Sheath Flow Chambers for Flow Cytometers : Micro Machined Flow Chamber with Low Pressure Loss
Ryo MIYAKEHiroshi OHKIIsao YAMAZAKITakeo TAKAGI
Author information
JOURNAL FREE ACCESS

1997 Volume 40 Issue 1 Pages 106-113

Details
Abstract
A new type of sheath flow chamber with low pressure loss was developed using micro-machining. The chamber was formed by laminating three 100-μm-thick, photoetched, metal plates. The device uses sheath-flow geometry : particles in suspension, which enter the chamber via an axial specimen nozzle, are enveloped by a buffering sheath flow and transported to a square capillary tube with cross-section of 300 μm×300μm. Finite element viscous-flow analysis is used to examine the flow behavior in the quasi-two-dimensional passage configuration which produces the fluid-dynamic focusing flow. Experimental measurement showed a smooth constricted sheath flow, and pressure loss was reduced to one-tenth of that which occurs with conventional glass flow chambers.