Abstract
This research investigated joint of the bone and the ligament in bone tunnel. The hole of 1mm diameter was made for the upper part of the tibia and the hole of 1mm, 1.5mm, and 2mm diameters was made under the tibia and the ligament was inserted. From the hole under the shinbone, the sample using three of the calcium phosphate paste, the alpha-cyanoacrylate adhesive was made, and the starch paste kinds of adhesive. Tensile tests at displacement rate of 1mm/s were performed. The alpha-cyanoacrylate adhesive force was significantly higher (p<0.01) than starch paste force. The calcium phosphate paste force was significantly higher (p<0.01) than starch paste force. The force increases in proportion to increase of amount of the injection of the adhesive.