The proceedings of the JSME annual meeting
Online ISSN : 2433-1325
2010.6
Session ID : J0204-2-5
Conference information
J0204-2-5 Evaluation in rotational stability for knee joint prosthesis
Kotaro IWAOHidehiko HIGAKITakeshi SHIMOTOYoshitaka SHIRAISHIHiromasa MIURAYukihide IWAMOTO
Author information
CONFERENCE PROCEEDINGS RESTRICTED ACCESS

Details
Abstract

This study investigated the contact pressure and rotation moment by loading of the femoral component and polyethylene insert in internal/external rotation. Five posterior stabilized prosthesis designs were tested to study using an artificial knee joint simulator. The contact pressure and rotational moment were measured at four flexion angle (0 degrees, 45 degrees, 90 degrees and 135 degrees) and five axial tibial rotations (0 degrees, 5 degrees, 10 degrees, 15 degrees and 20 degrees) under the applied loads of 1200N. With an internally rotated tibia, the contact pressure increased significantly and the rotation moment decreased significantly at high flexion. Additionally, Edge loading was found at the polyethylene insert at 20 degrees of tibial rotation. The design and stability with high adaptability in an excessive flexion rotation seem to be necessary.

Content from these authors
© 2010 The Japan Society of Mechanical Engineers
Previous article Next article
feedback