The Proceedings of the Symposium on Micro-Nano Science and Technology
Online ISSN : 2432-9495
2012.4
Session ID : P-OS3-9
Conference information
P-OS3-9 Development of calorimeter with MEMS thermopile sensor for biochemical analysis
Masataka SAITOJun TAKEUCHIOsamu NAKABEPPU
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Details
Abstract
Total biological activity measurement method through metabolic heat is referred as "Bio-calorimetry". We have developed batch-type and flow-type bio-calorimeter having MEMS thermopile sensors and multiple thermostat chambers. In the metabolic heat monitoring test of yeast cells using the batch-type, growth thermogram over several tens nW was measured. The batch-type has a weak point that signal cannot be obtained in the settling time of about 2 hours after beginning. Thus, fast response was required to the flow-type calorimeter as well as high sensitivity. Dilution tests of ethanol and NaCl aqueous solutions showed that the flow-type can monitor the thermal reaction immediately after the sample injection and has the settling time of about five minutes.