Journal of Japanese Society of Oral Implantology
Online ISSN : 2187-9117
Print ISSN : 0914-6695
ISSN-L : 0914-6695
Clinical Application of Endosseous Implant Treated by Electric Discharge Machining(IAT Fit II®
―Part 2. Abatment and Super-Structures―
Shigeru FujinoKazutaka SugiyamaToshio SagaraTakashi Miyazaki
Author information
JOURNAL FREE ACCESS

1999 Volume 12 Issue 2 Pages 238-245

Details
Abstract
In the previous report, endosseous implant system (IAT Fit II) treated by wire type electric discharge machining had excellent biocompatibility to bone and the high rate of achieving osseointegration in report 1. In this report, the longitudinal data of 217 IAT Fit II implants and 84 superstructures fixed in 64 patients between 1995 to 1997 is presented.
1. The survival rate of the implant bodies during the observation period was 99.5% with 100% of the superstructures remaining clinically functional.
2. Among the 84 superstructures fixed, 41 were cement-retained, and 43 were screw-retained. All superstructural components were bridge-type.
3. Complications occurred in 1 case with fixed-screws and 2 fractured implant bodies, which were replaced successfully.
4. No complication of loosing fixed-screw was reported during the observation period.
The results suggested that the ITA Fit II implant system implant body-abutment-superstructure-complex had excellent prognosis.
Content from these authors
© 1999 Japanese Society of Oral Implantology
Previous article Next article
feedback
Top