Journal of Japanese Society of Oral Implantology
Online ISSN : 2187-9117
Print ISSN : 0914-6695
ISSN-L : 0914-6695
A Studv on Fastening Torque Force, during Implant Reconstructive Procedure
Part 1. IMZ Implant
Yoshiyuki HagiwaraMasayuki KoizumiYoshinori NametaYoshio MoriyaMasaharu KamioTakayoshi Igarashi
Author information
JOURNAL FREE ACCESS

1993 Volume 6 Issue 2 Pages 370-373

Details
Abstract

The Osseointegrated Implant System was indispensable for many kinds of parts, such as transmucosal meso-structure or superstructure. They were fastened by a screw. The fastening force had an influence on the load of the bite force, accuracy of the superstructure, and prognosis of the implant. Therefore, in this study, we measured the optimum fastening torque force, by using a torque wrench.
The results obtained were as follows:
1. The clinical optimum fastening torque force for the impression post, IME, and superstructure was 4.5 Ncm.
2. The newly devised measuring system was easy to use and repeatable. It should be valuable for measuring the fastening force for various implant systems.

Content from these authors
© 1993 Japanese Society of Oral Implantology
Previous article Next article