Hyomen Kagaku
Online ISSN : 1881-4743
Print ISSN : 0388-5321
ISSN-L : 0388-5321
Originals
Development of a New High-resolution RBS System
Kenji KIMURAKaoru NAKAJIMA
Author information
JOURNAL FREE ACCESS

2001 Volume 22 Issue 7 Pages 431-437

Details
Abstract
A compact high-resolution RBS system consisting of a 90o sector magnetic spectrometer (radius 200 mm), a quadrupole lens, an electrostatic deflector and a one dimensional position sensitive detector (length 100 mm) has been developed. The measured energy resolution of the spectrometer is 0.11% of the analyzing energy at an acceptance angle of 0.3 msr including the energy spread of the incident ion beam. Some observations using this new system are presented.
Content from these authors

この記事はクリエイティブ・コモンズ [表示 - 非営利 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc/4.0/deed.ja
Previous article Next article
feedback
Top