Shinku
Online ISSN : 1880-9413
Print ISSN : 0559-8516
ISSN-L : 0559-8516
On the Effect of Thermo-Electrically Cooled Baffle in Oil Diffusion Pump System
Futoshi KANEMATSUHiroshi MIMAYoshiharu YOSHIKAWAHiroshi YAMAZAKI
Author information
JOURNAL FREE ACCESS

1968 Volume 11 Issue 3 Pages 88-92

Details
Abstract
The chevron baffle with the two-stage type thermo-module could be kept at -47°C when the hot side of the module was cooled to 11°C by water. The final pressure was reduced by about one order and reached ca. 1 × 10-8 Torr with this baffle. The main residual gas component was the vapor of diffusion pump oil, DC-704. When the baffle was cooled by water with-out thermo-electrical cooling, the vapor of rotary pump oil was predominant. Radio-frequency mass spectrometer was used for the residual gas analysis.
Content from these authors
© The Vacuum Society of Japan
Previous article Next article
feedback
Top