Long-term Quality Control Test of a 68Ge/68Ga Generator
Sho TomitaKei HigashikawaSatoshi UenoYuki MizunoShusaku TazawaYuji Kuge
Author information
JOURNAL OPEN ACCESS

2021 Volume 58 Issue 1 Pages 47-58

Details
Abstract

Quality of the 68Ga solution eluted from a 68Ge/68Ga generator (Galli Eo®) was evaluated. Elution was performed 488 times from 21 to 484 days (ca. 15 months) after the calibration date. The eluted 68Ga activity was 329 MBq to 1,148 MBq, which decreased with the decay of 68Ge, but no significant change was observed in the elution yields (62.8±2.2%). The half-life of the eluted radioactivity was 67.8±0.1 min, the radionuclide purity of 68Ga was ≥99.9%, the 68Ge breakthrough was 0.000024 ±0.000004%, and the radiochemical purity of 68Ga3+ was 99.7±0.2%. Fe and Zn were detected as metal impurities in the eluent, but both were ≤10 μg/GBq. The endotoxin concentration of the eluate was ≤5 EU/mL, and the eluent passed the sterility test. These results show that the generator can stably provide 68Ga solution over a 15-month period.