Studies on Phantom Test Procedures and Criteria for Standard FDG-PET Imaging
Ukihide TateishiKeiichi MatsumotoHiromitsu DaisakiKazufumi SuzukiNaoki ShimadaGo Akamatsu
Author information
JOURNAL OPEN ACCESS

2021 Volume 58 Issue 1 Pages 15-17

Details
Abstract

To establish a feasible and practical methodology for harmonization of FDG-PET, we have investigated quantitative physical performance of recent clinical PET scanners (23 scanners) using the NEMA image quality phantom. Compared with the previous phantom data (13 scanners) acquired in the early 2000s, the current phantom data showed better contrasts of small spheres even with a smaller number of coincidence counts. This result suggests that clinical PET image quality and small lesion detectability might have been significantly improved in the past 10 years. Based on the data acquired in this working group, we are planning to update the standard phantom test methodology and harmonizing criteria for FDG-PET.

Content from these authors
© 2021 Japanese Society of Nuclear Medicine

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 継承 4.0 国際]ライセンスの下に提供されています。