KENBIKYO
Online ISSN : 2434-2386
Print ISSN : 1349-0958
Researches Today
Development of Cc Corrector under the Triple C Project
Hidetaka SawadaFumio HosokawaTakeo SasakiShuuichi YuasaKoji KatazakaiShunsaku WakiMuneyuki KawazoeYuko ShimizuTomohiro NakamichiKazutomo Suenaga
Author information
JOURNAL FREE ACCESS

2015 Volume 50 Issue 1 Pages 71-75

Details
Abstract

The Cc corrector was developed by applying a combination concave-lens effect from a thick twofold astigmatism field. Magnetic and electrostatic twofold astigmatism fields produced by dodecapoles were used to introduce the concave lens effect. The Cc corrector was mounted below the objective lens, and the Delta-type corrector for a TEM was also installed below the Cc corrector in tandem to correct geometrical aberrations. The simultaneous Cc and Cs correction and atomic-resolution imaging on a TEM using the system were experimentally demonstrated at an accelerating voltage of 30 kV.

Content from these authors
© 2015 The Japanese Society of Microscopy
Previous article
feedback
Top