The Review of Laser Engineering
Online ISSN : 1349-6603
Print ISSN : 0387-0200
ISSN-L : 0387-0200
Laser Original
Basic Study on Synthesized Light Source for Optical Coherence Tomography
Manabu SATOIchiro WAKAKIKeiichi URUSHIYAMAYuuki WATANABENaohiro TANNO
Author information
JOURNAL FREE ACCESS

2003 Volume 31 Issue 10 Pages 663-667

Details
Abstract
We have studied on the light intensity ratio of synthesized light source (SLS) for the improvement of spatial resolution of optical coherence tomography (OCT). SLS consists of two LEDs with wavelength 691 nm, spectral width 99 nm and 882 nm, 76 nm. The small imaging interferometer also consists of plan-convex lens of focal length 6 mm and diameter 6 mm, beam splitter and reference mirror with PZT. The axial resolution was measured at 1.2 μm with the side lobe intensity 42 % on the condition of intensity ratio of 1:0.5. The irradiated power was 18 μW. This axial resolution was 57 % compared to the axial resolution using a single LED. The image of test pattern was measured using the phase shift method and two lines with the interval of 9.8 μm were clearly measured. The lateral resolution was calculated at 1.1 μm from wavelength and numerical aperture.
Content from these authors
© 2003 by The Laser Society of Japan
Previous article Next article
feedback
Top