The Review of Laser Engineering
Online ISSN : 1349-6603
Print ISSN : 0387-0200
ISSN-L : 0387-0200
Long Life Operation of a Copper Vapor Laser
Hironobu KIMURARyoichi OTANINobutada AOKIChikara KONAGAIShimpey SHIRAYAMATatuo MIYAZAWA
Author information
JOURNAL FREE ACCESS

1995 Volume 23 Issue 3 Pages 210-219

Details
Abstract
The operating life of a copper vapor laser (CVL) with a 6 cm diameter / 200 cm discharge length laser tube has been successfully extended and continuous operation for 320 hours at an output power of 75±10 W has been demonstrated. With a view to improving the long-term performance and reliability of CVL operation, this paper reports the experimental results of investigating the relationship between copper loss rate and buffer gas condition (flow rate and pressure). Also discussed are the optimum laser operating conditions for minimizing copper loss. Aveage thyratron life has also been extended to 1, 000 hours or more by using the magnetic assist technique for the CVL pulsed power supply.
Content from these authors
© The Laser Society of Japan
Previous article Next article
feedback
Top