Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
ORIGINAL PAPERS
Ion Optics of Single Focusing Mass Spectrometer
Hisashi Matsuda
Author information
JOURNAL FREE ACCESS

1981 Volume 29 Issue 2 Pages 161-166

Details
Abstract
Image aberrations of various different magnetic sector type mass spectrometers are calculated and compared. Four second order aberration coefficients Aαα, Ayy, A and Aββ are important. The third order aberration coefficients are sufficiently small. The old proposals for Aαα = 0 yield very large aberrations for Ayy, A and Aββ due to the influence of fringing field. The image aberrations of 90° and 60° sector type with symmetric arrangement are rather small. Several ion optical systems for a better single focusing mass spectrometer are proposed (for instance, φm = 130°, ε1 = ε2 = 55°, Aγm = 3).
Content from these authors
© 1981 by The Mass Spectrometry Society of Japan
Next article
feedback
Top