Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
REGULAR PAPERS
Computer-assisted Null Point Method for the Mass Spectrometric Measurement of Relative Abundance of Hydrogen Isotopes
Takashi OhsumiJun Tamura
Author information
JOURNAL FREE ACCESS

1986 Volume 34 Issue 4 Pages 197-206

Details
Abstract
A multiplying digital-to-analog converter (DAC) has been incorporated into the measuring circuit of a conventional double collector mass spectrometer. Inverse feedback of a portion of the signal from the mass-2 peak (H2+) is controlled by the microcomputer so as to exactly cancel the ion current signal of the mass-3 peak (HD+ + H3+). The control to obtain the null point was so successful that we can regard the digital number being sent to the DAC as the proportional value to the ion current ratio. The software simulation of gas exchange in the ionization chamber from the sample gas to the reference gas and vice versa brought about the reduction of the measurement time.
Content from these authors
© 1986 by The Mass Spectrometry Society of Japan
Next article
feedback