Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
REGULAR PAPERS
Analysis of Vulcanization Accelerator in Vulcanized Rubber by Thermal Desorption/Pyrolysis-GC/AED and MS
Satoshi ITO, Sadao NAKAMURA, Shigeki DAISHIMA, Chuichi WATANABE
Author information
JOURNAL FREE ACCESS

1998 Volume 46 Issue 4 Pages 336-341

Details
Abstract
In this work, a method for analysis of vulcanization accelerator in vulcanized rubber was developed by using a pyrolyzer (Py)-gas chromatograph/atomic emission detector (GC/AED) and gas chromatograph/mass spectrometer (GC/MS). Thermal desorption-GC/AED made it possible to monitor the sulfur compounds and amines decomposed from vulcanization accelerator under vulcanization. Specified peaks of AED chromatograms were identified by the corresponding MS data. Qualitative analysis of these compounds could estimate original chemical structures of the vulcanization accelerators.
Temperature programmed Py-AED made it possible to get evolved gas profile of each elements. Temperature programmed Py-MS could get evolved gas profile of each compounds from specific ion chromatograms.
Content from these authors
© 1998 by The Mass Spectrometry Society of Japan
Previous article Next article
feedback
Top