Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
NOTES
Pb Isotope Analyses Using Bead Method: Application to BCR-1 and GSJ Standard Rock Samples
Tomomi KANISusumu NOHDA
Author information
JOURNAL FREE ACCESS

2002 Volume 50 Issue 4 Pages 199-203

Details
Abstract
It has been established that a new filament loading scheme for lead isotopic analysis using thermal ionization mass spectrometry, so called “bead method,” which gives stable ion beam and high reproducibility. Two minor modifications to this scheme are presented in this paper; finer colloidal silicic acid is used instead of silica-gel, and flash heating is performed in nitrogen environment to prevent excess oxidation of filament. We report lead isotopic compositions of the NIST lead standard SRM 981, BCR-1, and GSJ standard rock samples using the present method.
Content from these authors
© 2002 by The Mass Spectrometry Society of Japan
Previous article Next article