Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
Studies on Data Processing of Instrumental Analysis by Small Computer(II).
Data Processing of GC-MS
SHOZO OSHIMATAKAO NISHISHITATADAO KONISHIYASUSHI SUGIYAMA
Author information
JOURNAL FREE ACCESS

1970 Volume 18 Issue 4 Pages 1254-1258

Details
Abstract
A data processing system of GS-MS having an excellent cost-performance was constructed by the introduction of the off-line system using an analog data recorder. HITACHI RMU-5B mass spectrometer was partially modified to process mass spectra, and the PDP-8/I digital computer with 8K core memory and the A/D converter was adopted. M/e values were calculated from the relations between mass number and time which is based on the He peak; then, there were no error to M/e300 during the operation of two monthes. Coefficient of variation of the pattern coefficient is about3to5%.