Journal of the Mass Spectrometry Society of Japan
Online ISSN : 1880-4225
Print ISSN : 1340-8097
ISSN-L : 1340-8097
The Analysis of 235U Isotope Ratio by A Mass Spectrometer for UF6
YUUKOU NAGATOROHIROMI TOKOIKATSUMICHI SUZUKIEISKE MITANINORIHIKO OZAKISEIJI TAKEDA
Author information
JOURNAL FREE ACCESS

1975 Volume 23 Issue 2 Pages 119-125

Details
Abstract
The mass spectrometer for the precise determination of 235isotope ratio on UF6has been developed.The resolving power of this apparatus is 330. The UF+5 ion currents are measured with the accuracy of less than 0.1% by controlling the room temperture and humidity and by shielding the preamplifier. The isotope ratios of 235U are determined with the relative error of less than 0.1% by using the interpolative method. The time required for measuring one sample is less than15minutes.
Content from these authors
© The Mass Spectrometry Society of Japan