Computer Program“TRIO”for Third Order Calculation of Ion Trajectory
TAKEKIYO MATSUOHISASHI MATSUDAYOSHITAKA FUJITAHERMANN WOLLNIK
Author information
JOURNAL FREE ACCESS

1976 Volume 24 Issue 1 Pages 19-62

Details
Abstract
A computer program for the calculation of ion trajectory is described. This program“TRIO”(Third Order Ion Optics)is applicable to any ion optical system consisting of drift spaces, cylindrical or toroidal electric sector fields, homogeneous or inhomogeneous magnetic sector fields, magnetic and electrostatic Q-lenses. The influence of the fringing field is taken into consideration. A special device is introduced to the method of matrix multiplication to shorten the calculation time and the required time proves to be about 40 times shorter than the ordinary method as a result. The trajectory calculation is possible to execute with accuracy up to third order. Any one of three dispersion bases, momentum, energy, mass and energy, is possible to be selected. Full LIST of the computer program and an example are given