Medical Imaging and Information Sciences
Online ISSN : 1880-4977
Print ISSN : 0910-1543
ISSN-L : 0910-1543
Evaluation of Radiographic Effects of X-Ray Tube-Voltage Ripple-Rates a
Masao MATSUMOTOHitoshi KANAMORI
Author information
JOURNAL FREE ACCESS

1987 Volume 4 Issue 3 Pages 95-100

Details
Abstract
An appreciable radiographic effect of the increase of a tube-voltage ripple-rate is the increase of a blur due to the motion of an object. This phenomenon is explained in terms of MTFs and information spectra. The advantage of three-phase units compared with single-phase units in the radiographic image quality is evaluated.
Content from these authors
© Medical Imaging and Information Sciences
Previous article Next article
feedback
Top