1991 Volume 8 Issue 3 Pages 86-94
Modulation transfer functions (MTFs) for the defocused objective lens system of the microdensitometer used for analysis of radiographic image qualities, are investigated for two objective lenses. Medical X-ray films have a unique structure characterized by the use of double-coated emu lsion. However, When MTFs are measured using an objective lens with a focus shallower than the thickness of the film in a screen-film system, large measurement errors do not occur. For an objective lens with a focal depth of 0.10mm, the measured MTFs errors are approximately 5% at 4 cycles/mm.
The Wiener spectrum values measured using lenses with focal depths of 0.1 0 m m and 0.24mm almost agree. However, when measured with the 0.24mm objective lens, the Wiener spectr a show significantly higher values in the low frequencies due to the effect of the stage glass.