Medical Imaging Technology
Online ISSN : 2185-3193
Print ISSN : 0288-450X
ISSN-L : 0288-450X
Papers
Correction of Echo Shift in Reconstruction Processing for Ultra-Short TE Pulse Sequence
Masahiro TAKIZAWATakehiro OOTSUKATakayuki ABETetsuhiko TAKAHASHI
Author information
JOURNAL FREE ACCESS

2010 Volume 28 Issue 5 Pages 343-350

Details
Abstract
An ultra-short TE pulse sequence is composed of a radial sampling that acquires echo signals radially in the K-space and a half-echo acquisition that acquires only half of the echo signal. The shift in the position of the echo signal (echo shift) caused by the timing errors in the gradient magnetic field pulses affects the image quality in the radial sampling with the half-echo acquisition. To improve image quality, we have developed a signal correction algorithm that detects and eliminates this echo shift during reconstruction by performing a pre-scan within 10 seconds. The results showed that image quality is improved under oblique and/or off-centering conditions that frequently cause image distortion due to hardware error. In conclusion, we have developed a robust ultra-short TE pulse sequence that allows wide latitude in the scan parameters, including oblique and off-centering conditions.
Content from these authors
© 2010 The Japanese Society of Medical Imaging Technology
Previous article Next article
feedback
Top