Abstract
If a kilo-voltage cone-beam computed tomography (CBCT) system mounted on a linear accelerator becomes available for dose calculation, we can confirm the dose distribution of treatment in each day by referring it to the initially planned dose distribution. In this paper, we verified the validity of the calculation method using CBCT images combined with multi-slice CT images. To evaluate the accuracy of calculated dose distribution, γ analysis, distance-to-agreement analysis and dose-volume-histogram analysis were used as the conventional dose calculation methods using CBCT images. The results showed that the dose distribution calculated by our proposed method agreed with the initial treatment plan better compared with the other methods. In addition, our method was so stable that the calculated dose distribution was insensitive to variations in clinical conditions. We demonstrated the feasibility of our proposed method for adaptive radiotherapy.